EPIPEN Autoloader
Notice Date
Product Description
EpiPen (epinephrine injection, USP) and EpiPen Jr (epinephrine injection, USP) Auto-Injectors
The recall impacts the 0.3 mg and 0.15 mg strengths of EpiPen Auto-Injector.Batch number 5FA665 Expiry Apr 17.
EpiPen Jr 2-Pak® Auto-Injectors, 0.15 mg 49502-501-02 5GN767 April 2017
EpiPen Jr 2-Pak® Auto-Injectors, 0.15 mg 49502-501-02 5GN773 April 2017
EpiPen 2-Pak® Auto-Injectors, 0.3 mg 49502-500-02 5GM631 April 2017
EpiPen 2-Pak® Auto-Injectors, 0.3 mg 49502-500-02 5GM640 May 2017
EpiPen Jr 2-Pak® Auto-Injectors, 0.15 mg 49502-501-02 6GN215 September 2017
EpiPen 2-Pak® Auto-Injectors, 0.3 mg 49502-500-02 6GM082 September 2017
EpiPen 2-Pak® Auto-Injectors, 0.3 mg 49502-500-02 6GM072 September 2017
EpiPen 2-Pak® Auto-Injectors, 0.3 mg 49502-500-02 6GM081 September 2017
EpiPen 2-Pak® Auto-Injectors, 0.3 mg 49502-500-02 6GM088 October 2017
EpiPen 2-Pak® Auto-Injectors, 0.3 mg 49502-500-02 6GM199 October 2017
EpiPen 2-Pak® Auto-Injectors, 0.3 mg 49502-500-02 6GM091 October 2017
EpiPen 2-Pak® Auto-Injectors, 0.3 mg 49502-500-02 6GM198 October 2017
EpiPen 2-pak® Auto-Injectors, 0.3 mg 49502-500-02 6GM087 October 2017
Defect
There is a potential that these devices may contain a defective part that may result in the device failing to activate or requiring increased force to activate.
Hazards
The failure of the auto-injector to activate may result in patients not receiving the required dose of adrenaline, causing the worsening of symptoms of anaphylaxis or anaphylactic reactions, which could be life threatening.
What To Do
Patients, customers and distributors are being notified and should refer to Mylan.com/EpiPenRecall for updates on product return and replacement instructions.
We are asking patients to keep their existing product until their replacement product can be secured.
Patients may receive either EpiPen Auto-Injector or the authorized generic for EpiPen Auto-Injector at the pharmacy as a replacement based on availability.
To return your product please contact Stericycle at 877-650-3494.
If you have any additional questions regarding this recall, please contact Mylan Customer Relations at 800-796-9526 or customer.service@mylan.com.
Where Sold
Worldwide
Traders Selling Product
Various locations.